The formula
is the average number of cells in one large square, the dilution factor and the concentration of your original suspension.
Where it comes from
A hemocytometer is a counting chamber with a grid of known area under a coverslip held at a known height, so every square holds a known volume.
On a standard Neubauer chamber each large square is 1 mm × 1 mm and the chamber is 0.1 mm deep:
One millilitre is one cubic centimetre, which is 1,000 mm³, so
Concentration is count divided by volume, so cells in the chamber per mL are
That is the concentration of what you loaded. If you diluted the cells first, for example 1:1 with trypan blue, the original was times more concentrated, giving .
Worked example
You count 42, 47, 44 and 47 cells in the four corner squares after mixing 1:1 with trypan blue:
Watch out for
- Count enough cells. Counting error scales with , so aim for at least 100 cells in total. That is about 10% error, and 400 cells is about 5%.
- Use one edge rule. Count cells touching the top and left lines but not the bottom and right, or you double count.
- Other chambers differ. This uses the Neubauer large square. Check your chamber’s square size and depth.